TEMP 2000A

(US Patent Number 5,659,397)

The AZ Technology TEMP 2000A Portable Emissometer represents substantial innovation and considerable advancement in laboratory portable instrumentation for determining ambient temperature total emittance. It is compact, light-weight, rugged, and ergonomically designed for ease of use in the field or in the laboratory.

The TEMP 2000A, by virtue of an advanced optical system, provides a portable reflectometer/emissometer that performs total hemispheric reflectance measurements from less than 3 to greater than 35 micrometers wavelength and is not limited in wavelength range due to filters, windows, or coatings, placed in the optical path. The TEMP 2000A optical system has been designed to minimize losses, to facilitate and maintain optical alignment, and to utilize the features of AZ Technology's patented ellipsoid collector that allows compactness for portability, efficiency, and measurement accuracy. The TEMP 2000A provides improved performance and maintainability over the Gier Dunkle DB-100 IR Reflectometer and in accordance with the ASTM E-408 standard.

The table below lists specifications for the TEMP 2000A.

Wavelength <3um to >35um (not limited by filters, windows, etc.)
Measurement accuracy (for specular and diffuse samples) - ± 1% of full scale for gray samples
- ± 3% of full scale for non-gray samples
Repeatability - ± 0.5% of full scale or better
Sample type Any Sample, including foils, insulators, etc.
Sample size and geometry -Flat Surfaces: ≥ 0.4 inches (1 cm) diameter
-
Concave Surfaces: 6.5 inches (16.5 cm) diameter
-
Convex Surfaces: 1 inch (2.5 cm) diameter
Sample temperature Room Temperature, Ambient
Displayed properties -Infrared Reflectance
-
Normal Emittance (300K)
-
Hemispheric emittance (300K)
Readouts

-Digital LCD panel meter
-Selectable IR emittance or reflectance display

Measurement range (reflectance) 0.00 to 1.00

Dimensions

-Optical Head: 5.25" diameter x 6.8" long
-Control and Display Unit: 4.5 x 7.75 x 7 inches
-Carry Case: 12.5 x 17 x 11 inches

Weight -Optical Head: 5 pounds
-Control and Display Unit: 4 pounds
-Carry Case: 11 pounds
Warranty 1 year parts and labor
 